M.M. Murzabekov, A.E. Eskin
FSUE “VNIIFTRI”, Mendeleevo, Moscow region, Russia
murzabekov@vniiftri.ru,
eskin@vniiftri.ru
Al’manac of Modern Metrology № 3 (31) 2022, pages 106–118
Abstract. Calibration of a fluxgate magnetometer is an integral part of its practical application. The well-known method of calibrating a magnetometer is based on its rotation around each of the axes. It requires a non-magnetic setup that allows rotation of the magnetometer in all three axes, and a constant external magnetic field during calibration. The article considers two algorithms for the magnetometer calibration method based on its rotation, presents the noise characteristics of two points chosen for magnetometer calibration, the calibration setup and the calibration results of three fluxgate magnetometers.
Keywords: fluxgate magnetometer, magnetometer calibration, Earth’s magnetic field.
Full texts of articles are available only in Russian in printed issues of the magazine.