A.E. Isaev1, Ghen Yi2, A.N. Matveev1, Jia Guanghui2, B.I. Khatamtaev1, Wang Shiquan2, N.G. Shcherblyuk1
1FSUE “VNIIFTRI”, Mendeleevo, Moscow region, Russia
2HAARI, Hangzhou, Fuyang Zhejiang, China
isaev@vniiftri.ru
Al’manac of Modern Metrology № 4 (32) 2022, pages 151–168
Abstract. The described COOMET project 786/RU-a/19: bilateral experimental comparisons of the calculated phase calibrations of hydrophones in the field at a frequency from 10 to 500 kHz between the Hangzhou Applied Acoustics Research Institute (HAARI) and the Russian Metrological Institute of Technical Physics and Radio Engineering (VNIIFTRI — pilot laboratory). The measurements were made, compiled and confirmed by comparing the uncertainty budgets of the measurement results of the module and the argument of the complex sensitivity of the reference hydrophone RHS-6. Typical results obtained in the frequency range of comparisons were 1.2–1.3 dB for the module and 0.14–0.16 radians for the hydrophone sensitivity index. The correctness of the methods used to evaluate the phase-phase calibrations of hydrophones in the field, including the method for calculating the frequency dependence of the indicator from the frequency dependence of the hydrophone sensitivity, has been confirmed.
Keywords: metrology of underwater sound measurements, COOMET pilot comparisons, amplitude-phase hydrophone calibration.
Full texts of articles are available only in Russian in printed issues of the magazine.